Two optical coherence tomography angiography (OCTA) analysis tools may demonstrate comparable reliability but substantially different absolute estimates in measuring macular vessel density in patients with glaucoma.
In the single-center study, researchers included 93 patients with open-angle glaucoma enrolled in the Individualised Perimetry Progression Observations in Glaucoma study. They analyzed macular OCTA images obtained during the first 2 study visits using the Erlangen-Angio (EA) tool and the OCTA vascular analyzer (OCTAVA). The visits occurred an average of 113 days apart, and 76 patients contributed images at both visits for the intersession reliability analysis.
The researchers compared vessel density across 4 macular sectors—superior, inferior, nasal, and temporal—and evaluated the repeatability of vessel density and foveal avascular zone (FAZ) area and circularity. Linear mixed-effects models assessed differences between tools, sectors, and visits, while repeatability coefficients, intraclass correlation coefficients, and coefficients of variation characterized measurement reliability. Regression analyses examined associations between vascular measurements and central visual field sensitivity after adjustment for image quality.
OCTAVA produced higher vessel density estimates than the EA tool, with a model-adjusted mean difference of 8.42 percentage points. At the first visit, average vessel density was 34% with OCTAVA vs 26% with the EA tool. Differences were observed across all 4 macular sectors, with the largest tool-dependent discrepancy identified in the inferior sector in the mixed-effects analysis.
Despite the differences, both tools demonstrated good repeatability for average vessel density. Intraclass correlation coefficients were 0.80 for the EA tool and 0.85 for OCTAVA. The corresponding repeatability coefficients were 5.36 and 5.97 percentage points, indicating that measurement differences of about 5 to 6 percentage points could occur between visits. The researchers reported comparable reliability between tools.
Sector-specific analyses showed that repeatability varied across macular regions. The inferior sector demonstrated the greatest measurement variability, with repeatability coefficients of 7.65 percentage points for the EA tool and 9.53 percentage points for OCTAVA. In the superior sector, the corresponding coefficients were 6.36 and 7.16 percentage points. OCTAVA showed larger repeatability coefficients across sectors, although the reliability estimates were comparable.
Higher average macular vessel density was associated with better central visual field sensitivity. After adjustment for image quality, regression coefficients were 0.38 for the EA tool and 0.57 for OCTAVA. The respective models explained 37% and 36% of the variation in vessel density, indicating similar associations between vascular measurements and visual function despite differences in the tools' absolute estimates.
FAZ area measurements demonstrated acceptable reliability, with intraclass correlation coefficients of 0.87 for the EA tool and 0.91 for OCTAVA. The tool produced larger FAZ area estimates, with a mean difference of 0.07 mm². In contrast, FAZ circularity demonstrated weaker reproducibility, particularly with OCTAVA, which had an intraclass correlation coefficient of 0.48 vs 0.71 for the EA tool. Neither FAZ area nor circularity showed a statistically significant association with central visual field sensitivity.
Differences in image-processing methods may contribute to the discrepancies. Methodologic differences may influence vessel identification and the resulting density estimates, limiting direct comparisons of absolute values across platforms.
The about 3 × 3-mm imaging area did not encompass all central visual field test locations, potentially affecting the observed associations between vessel density and visual function. The average 113-day interval between visits may have allowed genuine biological changes, influencing repeatability estimates. FAZ measurements were not corrected for ocular magnification because axial length data were unavailable. The single-center design and relatively small repeatability sample may also limit generalizability. The study did not directly establish whether either tool could reliably detect confirmed glaucoma progression.
“The quantitative differences between tools indicate that absolute values should not be compared across platforms, and the two tools cannot be used interchangeably for monitoring progression,” wrote lead study author Kowsigan Magesan, of the University of Western Australia, and colleagues.
Full disclosures of the study authors can be found in the study.
Source: Optometry and Vision Science