Objective:
Compare macular vessel-density and foveal avascular zone (FAZ) measurements from two OCTA analysis tools, including their reliability and associations with central visual field sensitivity in patients with open-angle glaucoma.
Approach:
- Study design: Single-center analysis of 93 participants in the Individualised Perimetry Progression Observations in Glaucoma study; 76 contributed images from both visits to the intersession reliability analysis.
- Measurements: Researchers analyzed macular OCTA images from two visits, averaging 113 days apart, with Erlangen-Angio (EA) and OCTAVA. They assessed vessel density in four sectors, FAZ area and circularity, repeatability, and associations with central visual field sensitivi…
Key Findings:
- OCTAVA estimated average vessel density at 34% at the first visit, compared with 26% for EA; the model-adjusted mean difference was 8.42 percentage points. Differences occurred in all four sectors, largest in the inferior sector.
- Both tools showed good average vessel-density repeatability: intraclass correlation coefficients were 0.80 for EA and 0.85 for OCTAVA. Repeatability coefficients were 5.36 and 5.97 percentage points, respectively.
- Higher average vessel density was associated with better central visual field sensitivity for both tools after adjustment for image quality; the models explained 37% of variation for EA and 36% for OCTAVA.
- FAZ area reliability was acceptable (intraclass correlation coefficients: 0.87 for EA and 0.91 for OCTAVA), while FAZ circularity was less reproducible, particularly with OCTAVA (0.48 vs 0.71 for EA). Neither FAZ measure was significantly associated with centr…
Interpretation:
The authors reported comparable reliability between tools but substantial differences in absolute measurements. They stated that absolute values should not be compared across platforms and that the tools cannot be used interchangeably for monitoring progression.
Limitations:
- The single-center design and relatively small repeatability sample may limit generalizability.
- The approximately 3 × 3-mm imaging area did not cover all central visual field test locations.
- The 113-day average between visits may have allowed biological changes that affected repeatability estimates.
- FAZ measurements were not corrected for ocular magnification because axial length data were unavailable.
- The study did not establish whether either tool could reliably detect confirmed glaucoma progression.